  MDS-DTM-RTD series battery powered benchtop meter optional e-mail: info@omega.com for latest product manuals: omegamanual.info shop online at omega.com ? users guide www.datasheet.net/ datasheet pdf - http://www..co.kr/
servicing north america: u.s.a.: omega engineering, inc., one omega drive, p.o. box 4047 iso 9001 certified stamford, ct 06907-0047 usa toll free: 1-800-826-6342 tel: (203) 359-1660 fax: (203) 359-7700 e-mail: info@omega.com canada: 976 berar laval (quebec), h7l 5a1, canada toll-free: 1-800-826-6342 tel: (514) 856-6928 fax: (514) 856-6886 e-mail: info@omega.ca for immediate technical or application assistance: u.s.a. and canada: sales service: 1-800-826-6342/1-800-tc-omega ? customer service: 1-800-622-2378/1-800-622-best ? engineering service: 1-800-872-9436/1-800-usa-when ? mexico/ tel: 001 (203) 359-1660 fax: 001 (203) 359-7700 latin america: e-mail: espanol@omega.com servicing asia: china: : 1698 yi shan road, unit 102 min hang district shanghai, china 201103 p.r.c. hotline: 800 819 0559/400 619 0559 e-mail: info@cn.omega.com servicing europe: benelux : toll-free: 0800 099 3344 tel: +31 20 347 21 21 fax: +31 20 643 46 43 e-mail: sales@omegaeng.nl czech republic: frystatska 184 733 01 karvin, czech republic tel: +420-59-6311899 fax: +420-59-6311114 e-mail: info@omegashop.cz france: toll-free: 0800 541 038 tel: 01 57 32 48 17 fax: 01 57 32 48 18 e-mail: esales@omega.fr germany/ austria: daimlerstrasse 26 d-75392 deckenpfronn, germany toll-free: 0800 8266342 tel: +49 (0) 7056 9398-0 fax: +49 (0) 7056 9398-29 e-mail: info@omega.de united kingdom: omega engineering ltd. iso 9001 certified one omega drive, river bend technology centre, northbank irlam, manchester m44 5bd united kingdom toll-free: 0800-488-488 tel: +44 (0) 161 777-6611 fax: +44 (0) 161 777-6622 e-mail: sales@omega.co.uk omeganet ? online service internet e-mail omega.com info@omega.com it is the policy of omega engineering, inc. to comply with all worldwide safety and emc/emi regulations that apply. omega is constantly pursuing certification of its products to the european new approach directives. omega will add the ce mark to every appropriate device upon certification. the information contained in this document is believed to be correct, but omega accepts no liability for any errors it contains, and reserves the right to alter specifications without notice. warning: these products are not designed for use in, and should not be used for, human applications. www.datasheet.net/ datasheet pdf - http://www..co.kr/
i MDS-DTM-RTD series benchtop meter table of contents page section 1 - introduction ....................................................................... 1-1 1.1 precautions ................................................................................................ 1-1 1.2 safety warnings and iec symbols ......................................................... 1-2 1.3 general description .................................................................................. 1-2 section 2 - hardware .......................................................................... 2-1 2.1 unpacking and inspection ...................................................................... 2-1 2.2 included items .......................................................................................... 2-1 2.3 dc power and battery connection ........................................................ 2-1 2.3.1 operation ..................................................................................... 2-1 2.3.2 power adaptor ............................................................................ 2-1 2.3.3 battery connection ..................................................................... 2-2 section 3 - installation and wiring ....................................................... 3-1 3.1 general meter dimensions ...................................................................... 3-1 3.2 front panel with overlay ......................................................................... 3-2 3.3 rear panel (terminal input, analog output, alarm output, wireless) 3-2 3.4 rear panel (m8 input, analog output, alarm output, wireless) ..... 3-3 3.5 rear panel (m12 input, analog output, alarm output, wireless) .... 3-3 3.6 rear panel (terminal input, analog output, alarm output) ............ 3-4 3.7 rear panel (m8 input, analog output, alarm output) ...................... 3-4 3.8 rear panel (m12 input, analog output, alarm output) .....................3-5 3.9 battery installation/replacement .......................................................... 3-6 3.10 installing wireless antenna .................................................................. 3-6 3.11 wiring (sensor, analog output, alarm output) ................................ 3-7 section 4 - setup & configuration ........................................................ 4-1 4.1 getting started .......................................................................................... 4-1 4.2 software utility ......................................................................................... 4-1 4.3 software installation ................................................................................ 4-1 4.4 configuration ............................................................................................ 4-4 section 5 - display features & meter operation .................................... 5-1 5.1 display features ....................................................................................... 5-1 5.2 keypad operation .................................................................................... 5-1 5.3 button operation ...................................................................................... 5-2 section 6 - optional wireless transmitter operation ............................ 6-1 6.1 introduction ............................................................................................... 6-1 6.2 rf communication basics ....................................................................... 6-1 6.3 basic system overview ............................................................................ 6-1 6.4 transmit rate vs. battery life ................................................................. 6-3 6.5 wireless transmitter setup ...................................................................... 6-5 section 7 - maintenance ...................................................................... 7-1 7.1 calibration ................................................................................................. 7-1 7.2 cleaning .................................................................................................... 7-1 7.3 fuse replacement ..................................................................................... 7-1 www.datasheet.net/ datasheet pdf - http://www..co.kr/
ii MDS-DTM-RTD series benchtop meter table of contents section 8 - specifications ..................................................................... 8-1 8.1 general .................................................................................................... 8-1 8.2 wireless option ......................................................................................... 8-2 section 9 - approvals, regulatory compliance ..................................... 9-1 9.1 fcc (domestic use) ................................................................................. 9-1 www.datasheet.net/ datasheet pdf - http://www..co.kr/
list of figures section figure description ............................................................... page section 1 1-1 iec symbols ................................................................1-2 table 1 MDS-DTM-RTD series benchtop meter versions .. 1-3 section 3 3-1 general meter dimensions ...................................... 3-1 3-2 front panel controls, indicators ............................. 3-2 3-3 rear panel connections (terminal input, analog output, alarm output, wireless) ............. 3-2 3-4 rear panel connections (m8 input, analog output, alarm output, wireless) ............. 3-3 3-5 rear panel connections (m12 input, analog output, alarm output, wireless) ............. 3-3 3-6 rear panel connections (terminal input, analog output, alarm output) .............................. 3-4 3-7 rear panel connections (m8 input, analog output, alarm output) .............................. 3-4 3-8 rear panel connections (m12 input, analog output, alarm output) ............................................. 3-5 3-9 battery installation/replacement .......................... 3-5 3-10 wireless antenna installation ................................. 3-6 3-11 wiring C temperature sensor .................................. 3-7 3-12 wiring C m8 connector ............................................ 3-7 3-13 wiring C m12 connector .......................................... 3-7 3-14 wiring C analog output .......................................... 3-8 3-15 wiring C alarm .......................................................... 3-8 section 4 4-1 software - welcome screen ...................................... 4-1 4-2 software C installation options screen .................. 4-2 4-3 software - select installation folder screen .......... 4-2 4-4 software C license agreement screen ................... 4-3 4-5 software C installation complete screen ............... 4-3 4-6 usb programming cable ......................................... 4-4 4-7 usb connector location .......................................... 4-4 4-8 launch setup utility screen .................................... 4-5 4-9 utility program - welcome screen .......................... 4-5 4-10 utility program C connect to digital gauge screen ........................... 4-6 4-11 utility program C verify connections screen ....... 4-6 4-12 utility program C establish link screen ................ 4-7 4-13 utility program C choose options screen ............. 4-8 4-14 analog output options screen ............................. 4-10 4-15 calibrations options screen C skip calibration option .......................................... 4-12 iii MDS-DTM-RTD series benchtop meter list of figures www.datasheet.net/ datasheet pdf - http://www..co.kr/
list of figures continued section figure description ............................................................... page 4-16 calibrations options screen C skip to next option ................................................ 4-12 4-17 send settings to digital gauge screen C finish option ........................................................... 4-13 section 5 5-1 display features ........................................................ 5-1 5-2 keypad operation ..................................................... 5-1 5-3 menu button operation ........................................... 5-3 5-4 front keypad set button .......................................... 5-5 section 6 6-1 fresnel zone ............................................................... 6-1 table 2 MDS-DTM-RTD battery life C wireless mode ..... 6-4 iv MDS-DTM-RTD series benchtop meter list of figures www.datasheet.net/ datasheet pdf - http://www..co.kr/
section 1 - introduction please read this manual completely before installing and operating your instrument. its important to read and follow all notes, cautions, warnings and safety precautions before setting up, installing and operating this unit. 1.1 precautions ? this device has not been designed, tested or approved for use in any medical or nuclear applications. ? never operate this device in flammable or explosive environments. ? never operate with a power source other than the one recommended in this manual. ? never operate this device outside of the recommended use outlined in this manual. ? follow all safety precautions and operating instructions outlined in this manual. ? keep unit out of reach of children. ? never operate with a power cord that is not properly rated for use with your unit. ? remove and/or disconnect main power cord before attempting any maintenance or fuse replacement. ? do not connect and/or operate this unit to a non-grounded, non-polarized outlet or power source. for models with wireless transmitter option ? no co-location with other radio transmitters is allowed. by definition, co- location is when another radio device or its antenna is located within 20 cm of your unit and can transmit simultaneously with your unit. ? never operate a wireless unit within 20 cm or less from each other. ? never install and/or continuously operate your wireless unit closer than 20 cm to nearby persons. there are no user serviceable parts inside your device. attempting to repair or service your unit may void your warranty. 1-1 introduction 1 note: www.datasheet.net/ datasheet pdf - http://www..co.kr/
1.2 safety warnings and iec symbols this device is marked with international safety and hazard symbols in accordance with en61010-1. it is important to read and follow all precautions and instructions in this manual before operating or commissioning this device as it contains important information relating to safety and emc. failure to follow all safety precautions may result in injury and/or damage to your calibrator. use of this device in a manner not specified by the manufacturer may impair protection devices and safety features provided by the unit. iec symbols description caution, risk of electric shock caution, refer to accompanying documentation figure 1-1. iec symbols all electrical connections and wiring should be performed by suitably trained personnel only. 1.3 general description omegas new MDS-DTM-RTD series benchtop meter has the functions of a digital temperature meter, enclosed in a benchtop case. it is ideal for laboratory use and applications requiring portable temperature measurement. there is one basic model, a single-channel unit that features a terminal, m8, or m12 input that is configured for use with a rtd signal. this meter is factory configured and calibrated. standard features include: internal battery or external dc power supply operation, analog output, and user programmable alarms. the optional built-in wireless transmitter allows for remote monitoring, chart recording and data logging. a variety of user configurable options and settings include: update rate, dampening, units, backlight level, alarm setting, analog output scaling, and auto-power-off. introduction 1 1-2 caution: www.datasheet.net/ datasheet pdf - http://www..co.kr/
omega offers six different versions of the MDS-DTM-RTD series benchtop meter. table 1: MDS-DTM-RTD series benchtop meter versions version terminal m8 connector m12 connector analog alarm wireless input input output output MDS-DTM-RTD MDS-DTM-RTD-w MDS-DTM-RTD-m8 MDS-DTM-RTD-m8-w MDS-DTM-RTD-m12 MDS-DTM-RTD-m12-w introduction 1 1-3 www.datasheet.net/ datasheet pdf - http://www..co.kr/
section 2 C hardware it is important that you read this manual completely and follow all safety precautions before operating this instrument. 2.1 unpacking & inspection remove the packing list and verify that you have received all your equipment. if you have any questions about the shipment, please call our customer service department at 1-800-622-2378 or 203-359-1660. we can also be reached on the internet at www.omega.com, e-mail: cservice@omega.com. when you receive the shipment, inspect the container and equipment for any signs of damage. note any evidence of rough handling in transit. immediately report any damage to the shipping agent. the carrier will not honor any damage claims unless all shipping material is saved for inspection. after examining and removing contents, save packing material and carton in the event reshipment is necessary. 2.2 included items the following items are supplied in the box. ? 1 MDS-DTM-RTD series benchtop meter ? 1 users guide ? 1 usb programming cable ? 1 setup utility software disk ? 2 d size alkaline batteries ? mating connector (m8 and m12 modles only) ? universal dc power adapter 2.3 dc power and battery connection 2.3.1 operation your benchtop meter will operate from 12 vdc to 24 vdc. 2.3.2 power adaptor a universal power adaptor has been included with your unit. dc power input cord universal adaptor 100-240 vac 2-1 hardware 2 note: www.datasheet.net/ datasheet pdf - http://www..co.kr/
hardware 2 2-2 all electrical connections and wiring should be performed by suitably trained personnel only. 2.3.3 battery connection your benchtop meter will also operate on 2 d cell alkaline batteries. follow instructions in section 3.9 for details on how to replace the batteries. note: www.datasheet.net/ datasheet pdf - http://www..co.kr/
3-1 installation, mounting & wiring 3 section 3 C installation & wiring 3.1 general meter dimensions figure 3-1. general meter dimensions 207 (8.13) 80 (3.13) dimensions mm (in) 93 (3.68) www.datasheet.net/ datasheet pdf - http://www..co.kr/
installation, mounting & wiring 3 3-2 3.2 front panel with overlay the front panels of all MDS-DTM-RTD versions are the same. the panels include the overlay, led screen, and keypad. figure 3-2. front panel controls, indicators 3.3 rear panel (terminal input, analog output, alarm output, wireless) figure 3-3. rear panel connections (terminal input, analog output, alarm output, wireless) f 0% 100% min max avg peak cal bat hal lal log set tx omega engineering, inc. stamford, ct. 06907-0047 MDS-DTM-RTD series meter caution: dispose of battery properly see manual for instructions battery: 2x d-cell, 1.8 vdc max 12 - 24 vdc fuse: f1a, 250v wireless antenna www.omega.com e-mail: info@omega.com 1 2 3 input 1 2 3 analog output 1 2 3 alarm output usb port installation category ii www.datasheet.net/ datasheet pdf - http://www..co.kr/
installation, mounting & wiring 3 3-3 3.4 rear panel (m8 input, analog output, alarm output, wireless) figure 3-4. rear panel connections (m8 input, analog output, alarm output, wireless 3.5 rear panel (m12 input, analog output, alarm output, wireless) figure 3-5. rear panel connections (m12 input, analog output, alarm output, wireless) omega engineering, inc. stamford, ct. 06907-0047 MDS-DTM-RTD series meter caution: dispose of battery properly see manual for instructions battery: 2x d-cell, 1.8 vdc max 12 - 24 vdc fuse: f1a, 250v wireless antenna www.omega.com e-mail: info@omega.com input 1 2 3 analog output 1 2 3 alarm output usb port installation category ii omega engineering, inc. stamford, ct. 06907-0047 MDS-DTM-RTD series meter caution: dispose of battery properly see manual for instructions battery: 2x d-cell, 1.8 vdc max 12 - 24 vdc fuse: f1a, 250v wireless antenna www.omega.com e-mail: info@omega.com input 1 2 3 analog output 1 2 3 alarm output usb port installation category ii www.datasheet.net/ datasheet pdf - http://www..co.kr/
installation, mounting & wiring 3 3-4 3.6 rear panel (terminal input, analog output, alarm output) figure 3-6. rear panel connections (terminal input, analog output, alarm output) 3.7 rear panel (m8 input, analog output, alarm output) figure 3-7. rear panel connections (m8 input, analog output, alarm output) omega engineering, inc. stamford, ct. 06907-0047 MDS-DTM-RTD series meter caution: dispose of battery properly see manual for instructions battery: 2x d-cell, 1.8 vdc max 12 - 24 vdc fuse: f1a, 250v antenna www.omega.com e-mail: info@omega.com 1 2 3 input 1 2 3 analog output 1 2 3 alarm output usb port installation category ii omega engineering, inc. stamford, ct. 06907-0047 MDS-DTM-RTD series meter caution: dispose of battery properly see manual for instructions battery: 2x d-cell, 1.8 vdc max 12 - 24 vdc fuse: f1a, 250v antenna www.omega.com e-mail: info@omega.com input 1 2 3 analog output 1 2 3 alarm output usb port installation category ii www.datasheet.net/ datasheet pdf - http://www..co.kr/
installation, mounting & wiring 3 3-5 3.8 rear panel (m12 input, analog output, alarm output) figure 3-8. rear panel connections (m12 input, analog output, alarm output) 3.9 battery installation/replacement to install or replace the battery in your end device you must first remove the cap on the battery holder located on the rear panel. this will allow you access to the batteries. figure 3-9. battery installation/ replacement lithium batteries may get hot, explode or ignite and cause serious injury if exposed to abusive conditions. be sure to follow the safety warnings listed below: ? do not use a different battery other than what is specified in this manual or product data sheet. ? do not discharge the battery using any device except your meter. ? do not place the battery in fire or heat the battery. ? do not store batteries with other hazardous or combustible materials. ? do not install the battery backwards so the polarity is reversed. omega engineering, inc. stamford, ct. 06907-0047 MDS-DTM-RTD series meter caution: dispose of battery properly see manual for instructions battery: 2x d-cell, 1.8 vdc max 12 - 24 vdc fuse: f1a, 250v antenna www.omega.com e-mail: info@omega.com input 1 2 3 analog output 1 2 3 alarm output usb port installation category ii + C C + www.datasheet.net/ datasheet pdf - http://www..co.kr/
installation, mounting & wiring 3 3-6 ? do not connect the positive terminal and negative terminal of the battery to each other with any metal object (such as wire). ? do not carry or store the battery together with metal objects. ? do not pierce the battery with nails, strike the battery with a hammer, step on the battery or otherwise subject it to strong impacts or shocks. ? do not solder directly onto the battery. ? do not expose battery to water or salt water, or allow the battery to get wet. ? do not disassemble or modify the battery. ? always insulate the terminals with adhesive tape or similar materials before disposal. ? immediately discontinue use of the battery if the battery emits an unusual smell, feels hot, changes color or shape, leaks or appears abnormal in any other way. ? do not place the battery in microwave ovens or high-pressure containers. 3.10 installing wireless antenna to install or replace the wireless antenna, screw the threaded portion of the antenna on the mounted wireless jack located on the rear panel of the end device. figure 3-10. wireless antenna installation wireless antenna wireless jack www.datasheet.net/ datasheet pdf - http://www..co.kr/
installation, mounting & wiring 3 3-7 3.11 wiring (sensor, analog output, alarm output) temperature sensor model MDS-DTM-RTD is designed to operate with a 3-wire, pt100 external sensor or integral probe assembly. wires are connected to tb1 located on the rear panel. figure 3-11. wiring - temperature sensor figure 3-12. wiring - m8 connector figure 3-13. wiring - m12 connector front view front view tb1 red wire black wires pt100 sensor 13 2 pt100 sensor red wire black wires m12 connector pt100 sensor red wire black wires 1 2 3 4 www.datasheet.net/ datasheet pdf - http://www..co.kr/
installation, mounting & wiring 3 3-8 analog output wiring example figure 3-14. wiring - analog output alarm wiring example driving a relay or low impedance input (open drain) figure 3-15. wiring - alarm C + + _ ground shield wire enclosure chart recorder j2 1 2 3 tb2 * 1n4004 diode alarm relay 200 ma max 24 vdc power supply C C + + 1 2 3 www.datasheet.net/ datasheet pdf - http://www..co.kr/
section 4 C setup & configuration 4.1 getting started this section outlines how to setup and configure your MDS-DTM-RTD benchtop meter before installation and use. all configuration settings are set and saved into your meter by connecting the included usb programming cable and running the software utility that was included with your unit on your computer. 4.2 software utility your computer should meet the following minimum requirements: ? pentium class processor ? hard drive space: 210 meg ? ram: 256 meg or higher ? 1 available usb port ? 1 cd-rom drive ? windows 2000, xp, vista (32bit) operating system or windows 7. 4.3 software installation insert the software cd that was included with your unit into the cd-rom drive on your pc. your system should begin the installation process automatically. figure 4-1. software - welcome screen this welcome screen will be visible on your computer. to continue installing the program click the next > button. 4-1 setup & configuration 4 www.datasheet.net/ datasheet pdf - http://www..co.kr/
figure 4-2. software - install options screen from this screen you select if you want the program icons to be placed on your desktop and to automatically install the usb drivers. to continue with installing the program click the next > button. figure 4-3. software - select installation folder screen from this screen you select the folder were you want the program files installed on your pc. the default setting will install the software under your program folders in a new folder named omega to continue with installing the program click the next > button. setup & configuration 4 4-2 www.datasheet.net/ datasheet pdf - http://www..co.kr/
figure 4-4. software - license agreement screen from this screen you must select agree to continue installing your program. after making your selection click the next > button. the setup wizard will now complete the process and install the software. figure 4-5. software - installation complete screen congratulations! you have just successfully installed the dg setup utility program on your pc. to end installing the program and close the setup wizard click the close button setup & configuration 4 4-3 www.datasheet.net/ datasheet pdf - http://www..co.kr/
4.4 configuration connecting your meter to your computer a usb programming cable was included with your unit. this cable is only used during the setup and configuration of your meter. this cable does not remain connected during normal use. figure 4-6. usb programming cable connect the a-type connector to your pc and then connect the b-type connector to the usb port on your meter. figure 4-7. usb connector location setup & configuration 4 4-4 usb note: note: www.datasheet.net/ datasheet pdf - http://www..co.kr/
setting up your meter now that you have connected your usb cable to your pc and to your meter, you can now complete the following steps to configure your meter before placing the unit into operation. you will be using the configuration software utility that you installed onto your pc. if you have not installed the configuration software utility you should do so now. step 1. launch setup utility program. to launch the setup utility program on your pc begin by finding and clicking on the dtg program icon that was placed on your computer desk top when you installed the software. figure 4-8. launch setup utility screen step 2. connecting & communicating with your meter figure 4-9. utility program - welcome screen after starting the setup utility program this will be the first screen you will see. click the next > button to proceed and continue setting up your meter. each screen will provide instruction details on how to proceed. setup & configuration 4 4-5 www.datasheet.net/ datasheet pdf - http://www..co.kr/
figure 4-10. utility program - connect to digital gauge screen if you have not already connected your meter to a usb port on your pc you must do this now before continuing. after your unit has been connected click the next > button to proceed and continue setting up your unit. figure 4-11. utility program - verify connection screen setup & configuration 4 4-6 www.datasheet.net/ datasheet pdf - http://www..co.kr/
figure 4-12. utility program - establish link screen after successful communication between your connector/transmitter has been established you can click the next > button to proceed and continue setting up your connector/transmitter. if you did not receive this confirmation of proper communication you should click the back button to try connecting again. setup & configuration 4 4-7 www.datasheet.net/ datasheet pdf - http://www..co.kr/
figure 4-13. utility program - choose options screen from this screen you will select the main operating settings for your end device. each end device must have a different address number for proper operation. after making your selections click the next > button to proceed and program your settings into your unit. (1) backlight options intensity here you can set how bright you want the backlighting to be when on. keep in mind that when used under battery power the led brightness has a direct effect on the life of the battery. keep to the lowest setting possible for your ambient light conditions to conserve of battery power. external power here you can set the amount of time you want the backlighting to stay on when activated and the unit is running on external power. you can only set the backlighting to be always on when the unit is powered by an external power supply. when set to 60 or 300 seconds the backlight will come on and then turn off after the allotted time has expired. setup & configuration 4 4-8 note: note: www.datasheet.net/ datasheet pdf - http://www..co.kr/
battery power here you can set the amount of time you want the backlighting to stay on when activated and the unit is running on battery power only. you can not set the backlighting to be always on when the unit is powered by battery power. when set to 10 or 30 seconds the backlight will come on and then turn off after the allotted time has expired. if backlighting is not required it is recommended that you select always off to preserve battery life. (2) sensor options sample interval here you can set how oftern the device samples the sensor and updates the lcd. the sample interval is always fixed at 0.382 seconds/sample (2.6 samples/second) when powered externally, regardless of the setting in the configuration wizard. the purpose of the sample interval setting is for conserving battery life, not for display damping or averaging, and therefore has no effect unless operating on battery power only. rtd alpha select the proper curve for the sensor or probe you will be measuring with your unit. if you ordered a unit with an integral probe, this setting will be factory locked to match the probe installed. (3) front panel options show decimal point select this option to turn on or off the last digit and decimal point disable bar graph select this option to remove the bar graph indicator from the display. (4) alarm output alarm on disabled - the alarm output is disabled and will not operate. rising - the alarm output activates only when the temperature meets or exceeds the high setpoint. falling - the alarm output activates only when the temperature meets or falls below the low setpoint. setup & configuration 4 4-9 note: note: note: www.datasheet.net/ datasheet pdf - http://www..co.kr/
rising & falling - the alarm output activates when either the temperature meets or exceeds the high setpoint or the temperature falls below the low setpoint hal (high alarm limit) setpoint set here the high value you want the alarm to activate at. hal (high alarm limit) deadband deadband is an area where no action occurs. the purpose is to prevent oscillation or repeated activation-deactivation cycles. if your process value will always be very close to your alarm setting you should adjust the deadband to be a few degrees different than your alarm value. lal (low alarm limit) setpoint set here the low value you want the alarm to activate at lal (low alarm limit) deadband deadband is an area where no action occurs. the purpose is to prevent oscillation or repeated activation-deactivation cycles. if your process value will always be very close to your alarm setting you should adjust the deadband to be a few degrees different than your alarm value. (5) analog output options mode select the type of analog output your application requires. you can leave the default setting if you will not be using the analog output feature. you must set the wire jumpers located on the pc board to match the analog output type you selected here in the setup process. figure 4-14. analog output options setup & configuration 4 4-10 note: analog output www.datasheet.net/ datasheet pdf - http://www..co.kr/
scaling here you can scale the analog output to correspond to the process reading value. 4-20 ma example: this table shows analog output values you should expect if you set 4ma = 0f and 20ma = 1000f process reading analog output value 0f 4.00 ma 250f 8.00 ma 500f 12.00 ma 750f 16.00 ma 1000f 20.00 ma 0-10 vdc example: this table shows analog output values you should expect if you set 0.0vdc = 0f and 10.0 vdc = 1000f process reading analog output value 0f 0.00 vdc 250f 2.50 vdc 500f 5.00 vdc 750f 7.50 vdc 1000f 10.00 vdc setup & configuration 4 4-11 www.datasheet.net/ datasheet pdf - http://www..co.kr/
figure 4-15. calibration options screen - skip calibration option from this screen you will select a calibration option. if the unit does not require calibration you should leave the default selected skip calibration and continue by clicking the next > figure 4-16. calibration options screen - skip to next option after making your selections click the next > setup & configuration 4 4-12 www.datasheet.net/ datasheet pdf - http://www..co.kr/
figure 4-17. send settings to digital gauge screen - finish option after your meter has been programmed click the finish button to close the utility program. setup & configuration 4 4-13 www.datasheet.net/ datasheet pdf - http://www..co.kr/
5-1 display features & meter operation 5 section 5 C display features & meter operation 5.1 display features figure 5-1. display features (1) process reading, (2) units, (3) bargraph, (4) status icons 5.2 keypad operation figure 5-2. keypad operations (1) mode button, (2) set button f 0% 100% min max avg peakcalbat hal lal logset tx 2 3 1 4 1 2 0% 100% min max avg peak cal bat hal lal log set tx www.datasheet.net/ datasheet pdf - http://www..co.kr/
display features & meter operation 5 5-2 5.3 button operation the mode and set buttons located of on the front display of your meter are activated by pushing down onto the keypad button symbol on the front meter label. mode button operation C to activate the mode button, push your finger down on the mode button and hold for three seconds until the unit enters the mode menu. then move away your finger. each time you push down on the mode button the meter will step through each parameter. set button operation - the set button has two primary functions. independently the button is used to turn the backlighting on. to activate the backlight, push your finger down on the set button and hold for three seconds until the backlighting turns on. the backlighting will remain on for the amount of time you selected during the setup and configuration process. if you selected always off when setting the meter up, then the set button will not turn the backlighting on from the front keypad. when in the mode menu, the set button is used for the following: 1. to reset the min, max, or avg values to the current temperature reading. all three are reset simultaneously, from any one of the three screens, min, max, or avg. 2. to silence the high or low alarm, when in the hal or lal screen and the alarm is configured for latching. this will cause the display to stop blinking, however, the alarm output will remain activated until the alarm condition clears. 3. to turn the wireless transmitter on or off while in the wrls screen. 4. to select units of measure while in the unit screen. . note: www.datasheet.net/ datasheet pdf - http://www..co.kr/
display features & meter operation 5 5-3 figure 5-3. menu button operation menu mode apply force 3 apply force then remove min, max and avg all set to value of current reading min, max and avg all set to value of current reading min, max and avg all set to value of current reading blinking stops alarm silenced blinking stops alarm silenced seconds then remove apply force then remove min value max value avg value battery value high alarm value low alarm value apply force then remove apply force then remove apply force then remove apply force then remove apply force then remove mode 0% 100% min 0% 100% min max 0% 100% max avg 0% 100% avg batt 0% 100% bat hal 0% 100% hal lal 0% 100% lal min value max value min 0% 100% max 0% 100% avg value avg 0% 100% high alarm value low alarm value hal 0% 100% hal lal 0% 100% lal apply force then remove apply force then remove apply force then remove apply force then remove apply force then remove continued next page apply force then remove wireless option wireless option wireless option apply force then remove apply force then remove wrls 0% 100% wrls 0% 100% wrls 0% 100% www.datasheet.net/ datasheet pdf - http://www..co.kr/
figure 5-3. menu button operation continued display features & meter operation 5 5-4 units units units units exit mode units continued from previous page apply force then remove apply force then remove apply force then remove apply force then remove apply force then remove apply force 3 seconds then remove unit 0% 100% f 0% 100% k 0% 100% 0% 100% exit 0% 100% r g 0% 100% www.datasheet.net/ datasheet pdf - http://www..co.kr/
display features & meter operation 5 5-5 backlighting operation the front keypad set button can be used to activate the backlighting feature. when activated, the backlighting will remain on for the period of time you selected during the setup and configuration in section 3. figure 5-4. front keypad set button apply force 3 seconds to activate backlight www.datasheet.net/ datasheet pdf - http://www..co.kr/
6-1 optional wireless transmitter operation 6 section 6 C optional wireless transmitter operation 6.1 introduction compared to a wired connection, a wireless meter provides much simpler installation. based on the physical principle of the propagation of radio waves, certain basic conditions should be observed. the following simple recommendations are provided to ensure proper installation and correct operation of your wireless meter. 6.2 rf communication basics your wireless transmitter sends wireless transmissions to a receiver. the receiver checks the incoming data for accuracy and processes this data for use by the measurement software on your pc. radio signals are electromagnetic waves, hence the signal becomes weaker the further it travels. while radio waves can penetrate some solid materials like a wall, they are dampened more than when a direct line-of-sight between the transmitting and receiving antenna exist. 6.3 basic system overview a basic wireless temperature system is comprised of only two main components; a signal conditioner with a built-in battery powered 2.4ghz radio transmitter, and a usb powered 2.4ghz radio receiver. installation when installing your meter it is important to position the unit in such a way as to optimize the antenna location within whats known as the fresnel zone. the fresnel zone can be thought of as a football-shaped invisible tunnel between two locations that provides a path for rf signals between your meter and your receiver. figure 6-1. fresnel zone fresnel zone receiver antenna transmitter antenna www.datasheet.net/ datasheet pdf - http://www..co.kr/
optional wireless transmitter operation 6 6-2 in order to achieve maximum range, the football-shaped path in which radio waves travel must be free of all obstructions. obstacles in the path (especially metal) will decrease the communication range between your meter and receiver. also, if the antennas are mounted just barely off the ground, over half of the fresnel zone ends up being obstructed by the earth resulting in significant reduction in range. to avoid this problem, the antennas should be mounted high enough off of the ground so that the earth does not interfere with the central diameter of the fresnel zone. it is important to understand that the environment may change over time due to new equipment or machinery being installed, building construction, etc. if new obstacles exist between your meter and receiver, the devices can be raised on one end or on both ends to clear the fresnel zone of obstructions. placing your meter in an application where the unit will be exposed to ambient temperatures above or below the operating limits specified in this manual will damage your unit and cause the unit to malfunction and produce incorrect operation. your meter has been shipped to you with a standard approved antenna. use of any other antenna than whats supplied with your meter will void all fcc and ic regulatory compliance. environment omegas wireless end devices and receiver units have been designed to be fixed mounted and operated in a clean and dry indoor environment. care should be taken to prevent the components of your wireless system from being exposed to moisture, toxic chemicals and extreme cold or hot temperature that are outside the specifications listed in this manual. the following is a list of basic good practice you should apply when operating your wireless system. ? never operate your wireless device or receiver outside the recommended environmental limits specified in this manual. ? never operate your wireless end device or receiver in flammable or explosive environments. note: note: caution: www.datasheet.net/ datasheet pdf - http://www..co.kr/
? never use your wireless end device or receiver in medical, nuclear or other dangerous applications where an interruption of readings can cause damage or harm. ? never operate your end device or receiver with any other battery or power source than whats specified in this manual or on the battery compartment label. ? no co-location with other radio transmitters is allowed. by definition, co- location is when another radio device or its antenna is located within 20 cm of your end device and can transmit simultaneously with your end device. ? never operate end devices within 20 cm or less from each other. ? never install and/or operate your end device closer than 20 cm to nearby persons. ? never operate your end device with any other antenna than what is supplied or listed here in this manual for approved use. 6.4 transmit rate vs. battery life many factors such as ambient temperature conditions and transmitting rate can have a big effect on the life of the battery used in your meter. transmitting data places a big demand of the battery in your unit. the transmit rate is the single most contributing factor in the life of the battery. the slower the transmit rate you set, the longer the battery in your device will last. the tables and graphs below give some estimates on how long the battery should last vs. the transmit rate you selected when you setup your meter under normal operating conditions. optional wireless transmitter operation 6 6-3 www.datasheet.net/ datasheet pdf - http://www..co.kr/
optional wireless transmitter operation 6 6-4 6.4.1 MDS-DTM-RTD wireless model - battery life table 2: MDS-DTM-RTD - wireless model - estimated battery life batteries: d cell alkaline battery 21 ah capacity analog seconds/ 0.5 5 15 30 sample seconds per wireless transmission: weeks weeks weeks weeks 2 69.5 106.8 111.3 112.4 3 80.1 133.9 141.0 142.8 5 91.1 168.0 179.3 182.3 10 101.6 207.7 225.1 229.9 15 105.7 225.4 246.1 251.9 30 110.1 246.5 271.4 278.4 45 111.6 254.4 281.0 288.6 60 112.4 258.5 286.1 293.9 120 113.6 265.0 294.0 302.3 rf off 114.9 271.9 302.5 311.2 www.datasheet.net/ datasheet pdf - http://www..co.kr/
optional wireless transmitter operation 6 6-5 6.5 wireless transmitter setup after connecting the usb cable and running the configuration software as outlined in section 3, you will complete the following steps to configure your wireless transmitter before placing the unit into operation. you will be using the configuration software utility that you installed onto your pc in section 3. if you have not installed the configuration software utility you should do so now. during this procedure you will be setting the following parameters in your transmitter. rf channel this setting determines the operating channel on which rf connections are made between the transmitter and receiver. the transmitter must be set to the same channel as the receiver in order for them to communicate. network id this sets the id of the network that the transmitter will be joining. it must match the setting of the receiver in order for them to communicate. receiver address this sets a unique address number for your receiver. later, when you set up your receiver you will again set the same number to receive readings from the corresponding transmitter unit(s). each receiver must be set for a different number for your system to operate correctly. if you will be using more than one receiver unit in your area it is important to set the transmitter address numbers to be a corresponding number in your tc-central software. see examples below. for the first receiver: set the address on your transmitters to 101, 102, 103, 104, etc. then set the address in your tc-central software to match. for the second receiver: set the channels on your transmitters to 201, 202, 203, 204, etc. then set the address in your tc-central software to match. this numbering scheme can be expanded to match the number of receivers you are using. gauge address this sets a unique address number into your transmitter. later, when you set up your measurement software you will use this address setting to receive readings from the corresponding unit(s). each transmitter must be set for a different address for your system to operate correctly. note: www.datasheet.net/ datasheet pdf - http://www..co.kr/
optional wireless transmitter operation 6 6-6 sample rate this will program your device to transmit 1 data reading to your receiver at a specified time interval. available settings are 2, 3, 4, 5, 10, 15, 20, 25, 30, 45, 60, 75, 90, 105 or 120 seconds. www.datasheet.net/ datasheet pdf - http://www..co.kr/
6-7 optional wireless transmitter operation 6 notes: www.datasheet.net/ datasheet pdf - http://www..co.kr/
section 7 C maintenance 7.1 calibration your MDS-DTM-RTD series benchtop meter has been built, tested, and factory calibrated to meet or exceed the specifications listed here in this manual. information is provided below on how to have your unit returned for service. if your meter requires service or factory re-calibration, please call our customer service department at 1-800-622-2378 or 203-359-1660. they will assist you in arranging the return of your meter. we can also be reached on the internet at www.omega.com, e-mail: cservice@omega.com remove all electrical connections and power before attempting any maintenance or cleaning. 7.2 cleaning lightly dampen a soft, clean cloth with a mild cleaning solution and gently clean the benchtop meter. 7.3 fuse replacement disconnect all power from source before attempting fuse replacement. for continued protection against the risk of fire, replace fuse with only the same size, type, rating, and safety approvals indicated here and on the rear panel of your unit. all power line connections to your unit are fused. when operating on single-phase power source such as 120 vac in the usa, the neutral line is also fused. for all models C12-24 vdc. main power fusing: 1 ea., f1a (rear panel) ul./csa/vde approval (5mm dia x 20mm long). 7-1 maintenance 7 note: caution: caution: warning: www.datasheet.net/ datasheet pdf - http://www..co.kr/
8-1 specifications 8 section 8 C specifications 8.1 general range: -51 to 538c (-60 to 1000f) accuracy: 0.2c (0.3f) or 0.1% of reading, whatever greater resolution: 0.1c (0.1f) operating environment: -40 to 70c (-40 to 158f); 5 to 100% rh non- condensing rtd probe/ sensor input: pt100, 3-wire display type: lcd with selectable backlight display digits: 4 digits, 9999 counts character height: 25.4 mm (1.0 in) computer interface: usb (one programming cable included) sample/ display rate: fixed at 2/sec battery power: two (2) d size alkaline batteries. recommended battery type: duracell type mn1300 d size batteries for best performance at low temperatures. external power: 12-24 vdc battery life (typical): 2 years (without backlighting use and wireless transmitter operation) analog output: user selectable 0 to 5 vdc, 0 to 10 vdc, or 4 to 20 ma; requires external 12 to 24 vdc power supply for operation enclosure: case C aluminum, handle C aluminum enclosure finish: case - painted, handle - anodized enclosure screw torque: 10 oz in. settings (usb/ software) auto-off: locked-on, 1 min, 2 min, 5 min, 10 min, 15 min, 30 min, 60 min units: c, f, k lock: allows for front button lock-out alarms: user selectable high and low alarm limits backlight: on/off, or 10 sec, 30 sec, 1 min, 5 min backlight color: normal operation C green, low alarm C blue, high alarm - red dimensions: 93.47 h x 206.50 w x 203.20 mm d (3.68 x 8.13 x 8.00"), height not including sensor calibrate: zero and span weight: 1.2 kg (2.65 ib) www.datasheet.net/ datasheet pdf - http://www..co.kr/
8.2 wireless option transmit sample rate: user programmable from 1 sample/2 min to 1 sample/every 2 sec radio frequency (rf) transceiver carrier: ism 2.4 ghz rf output power: 10 dbm (10 mw) range of rf link: outdoor line of sight: up to 120 m (400 ft) indoor/ urban: up to 40 m (130 ft) software (included free): requires windows ? 2000, xp or vista (32 bit) data transmitted to host: temperature reading, ambient temperature reading, rf transmit strength and battery level specifications 8 8-2 www.datasheet.net/ datasheet pdf - http://www..co.kr/
9-1 approval, regulatory compliance 9 section 9 C approvals, regulatory compliance all approvals outlined in this manual are based on testing that was done with antennas that are supplied with your meter. removing and or installing a different antenna will void the product compliance demonstrated in these documents. 9.1 fcc (domestic use) for united states: fcc id: our-xbeepro for canada: ic #4214a-xbeepro this device complies with part 15 of the fcc rules. operation is subject to the following two conditions: 1.) this device may not cause harmful interference. 2.) this device must accept any interference received, including interference that may cause undesired operation. to satisfy fcc rf exposure requirements for mobile transmitting devices, a separation distance of 20 cm or more should be maintained between the antenna of this device and persons during device operation. to ensure compliance, operations at closer than this distance is not recommended. the antenna used for this transmitter must not be co-located in conjunction with any other antenna or transmitter. it is your (the users) responsibility to ensure that these products are operated within the guidelines here in this manual and in conformance with all local, state, federal or national regulations and laws of the country they are being operated in. transmitting power - your wireless series system components have been designed, manufactured and tested so that the transmitting power of your wireless meter will not exceed 10 dbm. note: note: note: warning: www.datasheet.net/ datasheet pdf - http://www..co.kr/
warrant y/ disclaimer omega engineering, inc. warrants this unit to be free of defects in materials and workmanship for a period of 25 months from date of purchase. omegas warranty adds an additional one (1) month grace period to the normal two (2) year product warranty to cover handling and shipping time. this ensures that omegas customers receive maximum coverage on each product. if the unit malfunctions, it must be returned to the factory for evaluation. omegas customer service department will issue an authorized return (ar) number immediately upon phone or written request. upon examination by omega, if the unit is found to be defective, it will be repaired or replaced at no charge. omegas warranty does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. this warranty is void if the unit shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of omegas control. components in which wear is not warranted, include but are not limited to contact points, fuses, and triacs. omega is pleased to offer suggestions on the use of its various products. however, omega neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by omega, either verbal or written. omega warrants only that the parts manufactured by the company will be as specified and free of defects. omega makes no other warranties or representations of any kind whatsoever, expressed or implied, except that of title, and all implied warranties including any warranty of merchantability and fitness for a particular purpose are hereby disclaimed. limitation of liability: the remedies of purchaser set forth herein are exclusive, and the total liability of omega with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. in no event shall omega be liable for consequential, incidental or special damages. conditions: equipment sold by omega is not intended to be used, nor shall it be used: (1) as a basic component under 10 cfr 21 (nrc), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. should any product(s) be used in or with any nuclear installation or activity, medical application, used on humans, or misused in any way, omega assumes no responsibility as set forth in our basic warranty/disclaimer language, and, additionally, purchaser will indemnify omega and hold omega harmless from any liability or damage whatsoever arising out of the use of the product(s) in such a manner. return requests/inquiries direct all warranty and repair requests/inquiries to the omega customer service department. before returning any product(s) to omega, purchaser must obtain an authorized return (ar) number from omegas customer service department (in order to avoid processing delays). the assigned ar number should then be marked on the outside of the return package and on any correspondence. the purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in transit. for warranty returns, please have the following information available before contacting omega: 1. purchase order number under which the product was purchased, 2. model and serial number of the product under warranty, and 3. repair instructions and/or specific problems relative to the product. for non-warranty repairs, consult omega for current repair charges. have the following information available before contacting omega: 1. purchase order number to cover the cost of the repair, 2. model and serial number of the product, and 3. repair instructions and/or specific problems relative to the product. omegas policy is to make running changes, not model changes, whenever an improvement is possible. this affords our customers the latest in technology and engineering. omega is a registered trademark of omega engineering, inc. ? copyright 2012 omega engineering, inc. all rights reserved. this document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without the prior written consent of omega engineering, inc. www.datasheet.net/ datasheet pdf - http://www..co.kr/
m5024/1012 where do i find everything i need for process measurement and control? om egaof course! shop online at omega.com sm temperature  thermocouple, rtd & thermistor probes, connectors, panels & assemblies  wire: thermocouple, rtd & thermistor  calibrators & ice point references  recorders, controllers & process monitors  infrared pyrometers pressure, strain and force  transducers & strain gages  load cells & pressure gages  displacement transducers  instrumentation & accessories flow/level  rotameters, gas mass flowmeters & flow computers  air velocity indicators  turbine/paddlewheel systems  totalizers & batch controllers ph/conductivity  ph electrodes, testers & accessories  benchtop/laboratory meters  controllers, calibrators, simulators & pumps  industrial ph & conductivity equipment data acquisition  data acquisition & engineering software  communications-based acquisition systems  plug-in cards for apple, ibm & compatibles  data logging systems  recorders, printers & plotters heaters  heating cable  cartridge & strip heaters  immersion & band heaters  flexible heaters  laboratory heaters environmental monitoring and control  metering & control instrumentation  refractometers  pumps & tubing  air, soil & water monitors  industrial water & wastewater treatment  ph, conductivity & dissolved oxygen instruments www.datasheet.net/ datasheet pdf - http://www..co.kr/


▲Up To Search▲   

 
Price & Availability of MDS-DTM-RTD

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X